MRI in Patients with PPM and ICD

Russo et al report on safety of conducting nonthoracic MRI in patients with non-conditional permanent pacemakers or implantable cardioverter-defibrillators. The protocol and supplement provide useful information.

Abstract

BACKGROUND

The presence of a cardiovascular implantable electronic device has long been a contraindication for the performance of magnetic resonance imaging (MRI). We established a prospective registry to determine the risks associated with MRI at a magnetic field strength of 1.5 tesla for patients who had a pacemaker or implantable cardioverter– defibrillator (ICD) that was “non–MRI-conditional” (i.e., not approved by the Food and Drug Administration for MRI scanning).

METHODS

Patients in the registry were referred for clinically indicated nonthoracic MRI at a field strength of 1.5 tesla. Devices were interrogated before and after MRI with the use of a standardized protocol and were appropriately reprogrammed before the scanning. The primary end points were death, generator or lead failure, induced arrhythmia, loss of capture, or electrical reset during the scanning. The secondary end points were changes in device settings.

RESULTS

MRI was performed in 1000 cases in which patients had a pacemaker and in 500
cases in which patients had an ICD. No deaths, lead failures, losses of capture, or ventricular arrhythmias occurred during MRI. One ICD generator could not be interrogated after MRI and required immediate replacement; the device had not been appropriately programmed per protocol before the MRI. We observed six cases of self-terminating atrial fibrillation or flutter and six cases of partial electrical reset. Changes in lead impedance, pacing threshold, battery voltage, and P-wave and R-wave amplitude exceeded prespecified thresholds in a small number of cases. Repeat MRI was not associated with an increase in adverse events.

CONCLUSIONS

In this study, device or lead failure did not occur in any patient with a non–MRIconditional pacemaker or ICD who underwent clinically indicated nonthoracic
MRI at 1.5 tesla, was appropriately screened, and had the device reprogrammed in
accordance with the prespecified protocol.

 

mri01mri01bmri02mri03

Advertisements
Categories: Uncategorized Tags: , ,
  1. No comments yet.
  1. No trackbacks yet.

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out / Change )

Twitter picture

You are commenting using your Twitter account. Log Out / Change )

Facebook photo

You are commenting using your Facebook account. Log Out / Change )

Google+ photo

You are commenting using your Google+ account. Log Out / Change )

Connecting to %s

%d bloggers like this: