Bioresorbable Vascular Scaffold Thrombosis

Wykrzykowska et al (N Engl J Med 2017;376:2319-28. DOI: 10.1056/NEJMoa1614954) report on a randomized clinical trial of bioresorbable vascular scaffold versus metallic drug eluting stent. The rates of target vessel failure were similar. However, scaffold thrombosis was significantly higher than stent thrombosis. There is an instructive accompanying editorial (DOI: 10.1056/NEJMe1703202).

ABSTRACT

BACKGROUND

Bioresorbable vascular scaffolds were developed to overcome the shortcomings of drugeluting stents in percutaneous coronary intervention (PCI). We performed an investigatorinitiated,randomized trial to compare an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent in the context of routine clinical practice.

METHODS

We randomly assigned 1845 patients undergoing PCI to receive either a bioresorbable vascular scaffold (924 patients) or a metallic stent (921 patients). The primary end point was target-vessel failure (a composite of cardiac death, target-vessel myocardial infarction, or target-vessel revascularization). The data and safety monitoring board recommended early reporting of the study results because of safety concerns. This report provides descriptive information on end-point events.

RESULTS

The median follow-up was 707 days. Target-vessel failure occurred in 105 patients in the scaffold group and in 94 patients in the stent group (2-year cumulative event rates, 11.7% and 10.7%, respectively; hazard ratio, 1.12; 95% confidence interval [CI], 0.85 to 1.48; P = 0.43); event rates were based on Kaplan–Meier estimates in time-to-event analyses. Cardiac death occurred in 18 patients in the scaffold group and in 23 patients in the stent group (2-year cumulative event rates, 2.0% and 2.7%, respectively), target-vessel myocardial
infarction occurred in 48 patients in the scaffold group and in 30 patients in the stent group (2-year cumulative event rates, 5.5% and 3.2%), and target-vessel revascularization occurred in 76 patients in the scaffold group and in 65 patients in the stent group (2-year cumulative event rates, 8.7% and 7.5%). Definite or probable device thrombosis occurred in 31 patients in the scaffold group as compared with 8 patients in the stent group (2-year cumulative event rates, 3.5% vs. 0.9%; hazard ratio, 3.87; 95% CI, 1.78 to 8.42; P<0.001).

CONCLUSIONS

In this preliminary report of a trial involving patients undergoing PCI, there was no significant difference in the rate of target-vessel failure between the patients who received a bioresorbable scaffold and the patients who received a metallic stent. The bioresorbable scaffold was associated with a higher incidence of device thrombosis than the metallic stent through 2 years of follow-up.

scaff02scaff01

Advertisements

Scaffold Thrombosis in Routine PCI

Wykrzykowska et al report a randomized clinical trial of metallic stent v bioresorbable vascular scaffold in routine percutaneous coronary intervention. They found significantly higher risk of device thrombosis with BVS. The editorial is available here.

Summary

BACKGROUND

Bioresorbable vascular scaffolds were developed to overcome the shortcomings of drugeluting stents in percutaneous coronary intervention (PCI). We performed an investigator-initiated, randomized trial to compare an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent in the context of routine clinical practice.

METHODS

We randomly assigned 1845 patients undergoing PCI to receive either a bioresorbable vascular scaffold (924 patients) or a metallic stent (921 patients). The primary end point was target-vessel failure (a composite of cardiac death, target-vessel myocardial infarction, or target-vessel revascularization). The data and safety monitoring board recommended early reporting of the study results because of safety concerns. This report provides descriptive information on end-point events.

RESULTS

The median follow-up was 707 days. Target-vessel failure occurred in 105 patients in the scaffold group and in 94 patients in the stent group (2-year cumulative event rates,11.7% and 10.7%, respectively; hazard ratio, 1.12; 95% confidence interval [CI], 0.85 to 1.48; P = 0.43); event rates were based on Kaplan–Meier estimates in time-to-event analyses. Cardiac death occurred in 18 patients in the scaffold group and in 23 patients in the stent group (2-year cumulative event rates, 2.0% and 2.7%, respectively), target-vessel myocardial infarction occurred in 48 patients in the scaffold group and in 30 patients in the stent group (2-year cumulative event rates, 5.5% and 3.2%), and target-vessel revascularization occurred in 76 patients in the scaffold group and
in 65 patients in the stent group (2-year cumulative event rates, 8.7% and 7.5%). Definite or probable device thrombosis occurred in 31 patients in the scaffold group as compared with 8 patients in the stent group (2-year cumulative event rates, 3.5% vs. 0.9%; hazard ratio, 3.87; 95% CI, 1.78 to 8.42; P<0.001).

CONCLUSIONS

In this preliminary report of a trial involving patients undergoing PCI, there was no significant difference in the rate of target-vessel failure between the patients who received a bioresorbable scaffold and the patients who received a metallic stent. The bioresorbable scaffold was associated with a higher incidence of device thrombosis than the metallic stent through 2 years of follow-up.

 

bvst01bvst02

Very Late Scaffold Thrombosis of Bioresorbable Vascular Scaffold

bvsvees.png
Toyota et al report a systematic review and meta-analysis of scaffold/stent thrombosis bioresorbable vascular scaffold versus everolimus eluting stent. The editorial is available here.

Abstract

OBJECTIVES

This study sought to compare the 2-year outcomes between bioresorbable vascular scaffold (BVS) and everolimus-eluting metallic drug-eluting stent (EES).

BACKGROUND

The occurrence of very late stent/scaffold thrombosis (VLST) of BVS beyond 1 year after implantation is an increasing concern.

METHODS

We conducted a meta-analysis of 24 studies (BVS: n ¼ 2,567 and EES: n ¼ 19,806) reporting the 2-year outcomes of BVS and/or EES to compare the risk of BVS versus EES for stent/scaffold thrombosis (ST) and target lesion failure (TLF) in 7 comparative studies (3 randomized and 4 observational), and to estimate the pooled incidence rates of ST and TLF including additional 17 single-arm studies.

RESULTS

In the 7 comparative studies, the risk for VLST between 1 and 2 years was numerically higher in BVS than in EES (odds ratio [OR]: 2.03 [95% confidence interval (CI): 0.62 to 6.71]). The excess risk of BVS relative to EES for ST
through 2 years was significant (OR: 2.08 [95% CI: 1.02 to 4.26]). The risk for TLF was neutral between BVS and EES. In the 24 studies, the pooled estimated incidence rates of VLST, and ST through 2 years were higher in BVS than in EES
(0.240 [95% CI: 0.022 to 0.608]% vs. 0.003 [95% CI: 0.000 to 0.028]%, and 1.43 [95% CI: 0.67 to 2.41]% vs. 0.56[95% CI: 0.43 to 0.70]%, respectively). The corresponding rates for TLF were comparable between BVS and EES(1.88 [95% CI: 1.30 to 2.55]% and 1.78 [95% CI: 1.17 to 2.49]% and 7.90 [95% CI: 6.26 to 9.69]% and 7.49 [95% CI:5.86 to 9.29]%, respectively).

CONCLUSIONS

In this meta-analysis, BVS as compared with EES was associated with higher risk for VLST between 1 and 2 years and ST through 2 years.